Bioelectrical Impedance Analysis Measures the Ejection Fraction of the Calf Muscle Pump
- Jan 01, 2007
- L.c Ward + 3 more +3
Chronic venous ulceration has been associated with dysfunction of the calf muscle pump (CMP). CMP ejection fraction can be used as an index of pump activity and is typically measured using air or strain gauge plethysmography (APG and SPG respectively) but both methods are technically complex and frequently exhibit poor reproducibility. In this study the ejection fraction of the CMP was assessed using bioelectrical impedance analysis (BIA). Calf resistance was recorded continuously at discrete frequencies (range 5 – 1000 kHz) using an Impedimed SFB7. Sensing electrodes were placed 20 cm apart over the calf muscle of the leg with the proximal electrode 7 cm distal to the crease at the back of the knee. Current input electrodes were placed at the base of the toes and fingers. Recordings were obtained from subjects undertaking the standard sequence of postural changes used to assess CMP activity: supine position with leg elevated (10 min); standing (5 min); plantar flexion of the ankle (10 repetitions in 10 seconds). Resistances representing residual volume and the functional venous volumes were determined from the resistance-time plots to calculate.