- Peer Review Report
- 10.5194/amt-2024-61-rc2
Comment on amt-2024-61
- Jun 12, 2024
- Steven A Bailey + 3 more +3
The Portable (2.7 kg) Cavity-enhanced Absorption of Nitrogen Dioxide (PCAND) instrument for measuring in situ nitrogen dioxide (NO2) was developed using incoherent, broadband cavityenhanced absorption spectroscopy (IBBCEAS). An LED light source centered at 408 nm was coupled to a cavity 15 cm in length, achieving an effective optical pathlength of ~520 m. Our precision was measured as 94 pptv (1 s). To date, we have flown this instrument on 3 balloon and 1 UAV test flights. This instrument records data to an SD card and outputs data (via an RS232 port) to external devices including a commercial radiosonde (iMet) for real-time data downlink. Nitrogen dioxide (NO2) is a major contributor to air pollution in the Earth's troposphere. Its main source is a byproduct of combustion from the burning of fossil fuels (Spinei, E. et al. 2014 ). NO2 has been monitored from satellite instruments (like OMI, TROPOMI, and GEMS) for a decade
Read more