4D attenuation map generation in PET/MR imaging using 4D PET derived motion fields

  • Oct 1, 2013
  • Hadi Fayad +3 more
Show More
  • Abstract
  • Literature Map
  • References
  • Citations
  • Similar Papers
Abstract

Respiratory motion can potentially reduce accuracy in functional and anatomical images fusion from combined PET/MR systems. Methodologies for the correction of respiratory motion in PET acquisitions in combined PET/MR systems are almost exclusively based on the use of respiratory synchronized MRI acquisitions to derive motion fields. The acquisition of 4D MR datasets is however associated with a compromise between image quality and acquisition duration limiting its use in clinical practice. The objective of this work was therefore to generate 4D MR images and associated attenuation maps from a single static MR image and motion fields obtained from simultaneously acquired 4D non-attenuation corrected (NAC) PET images. 4D PET/MRI datasets were acquired for five patients on the SIEMENS mMR PET/MR system. The 4D PET acquired datasets were retrospectively binned into 4 motion amplitude frames corresponding to the simultaneously acquired 4D gradient echo MR images. A single MR Dixon scan was also acquired at end of expiration for PET AC purposes. All the reconstructed 4D NAC PET images were then registered using an elastic registration algorithm to the single `end of expiration' NAC PET image. The derived motion fields were subsequently applied to the `end expiration' 3D gradient echo MR volume and its corresponding attenuation map to generate respiratory synchronized MR images and corresponding attenuation maps. The accuracy of the proposed method was assessed by comparing the generated to the original images. Good diaphragm profile matching, high correlation coefficients (0.91± 0.03) and small errors (<;3mm) were noticed between the generated and acquired 4D MRI series. Moreover, small classification errors between the generated and 4D MR extracted attenuation maps. As a conclusion, these results may allow for accurate PET AC and respiratory motion correction in PET/MR, without the need for patient specific 4D MR acquisitions.

Similar Papers
  • Conference Article
  • Citations5

Cardiac gating of transmission data is unnecessary for attenuation compensation of double-gated emission scans

  • Nov 08, 1998
  • G Klein +3
  • Conference Article
  • Citations1

Mismatch Correction for Free-breathing PET and Deep-inspiration Breath-holding CT in PET/CT Imaging

  • Oct 01, 2017
  • Hongcheng Shi +4
  • Conference Article
  • Citations3

Application of Conditional Adversarial Networks for Automatic Generation of MR-based Attenuation Map in PET/MR

  • Nov 01, 2018
  • Li Tao +3
  • Research Article
  • Citations3

Correlation between 2D and 3D Fetal Brain MRI Biometry and Neurodevelopmental Outcomes in Fetuses with Suspected Microcephaly and Macrocephaly.

  • Aug 12, 2021
  • American Journal of Neuroradiology
  • S Fried +7
  • Research Article
  • Citations37

Relative role of motion and PSF compensation in whole‐body oncologic PET‐MR imaging

  • Mar 24, 2014
  • Medical Physics
  • Yoann Petibon +7
  • Research Article
  • Citations2

An Investigation on Cross-Tracer Generalizability of Deep Learning-based PET Attenuation Correction.

  • Nov 01, 2025
  • IEEE transactions on radiation and plasma medical sciences
  • Jun Hou +10
  • Research Article
  • Citations55

Reconstruction-Incorporated Respiratory Motion Correction in Clinical Simultaneous PET/MR Imaging for Oncology Applications.

  • Apr 23, 2015
  • Journal of Nuclear Medicine
  • Hadi Fayad +3
  • Research Article
  • Citations40

Non-rigid dual respiratory and cardiac motion correction methods after, during, and before image reconstruction for 4D cardiac PET

  • Dec 01, 2015
  • Physics in Medicine & Biology
  • Tao Feng +3
  • Research Article
  • Citations6

PET/CT-guided biopsy with respiratory motion correction

  • Sep 11, 2019
  • International Journal of Computer Assisted Radiology and Surgery
  • Ruoqiao Zhang +7
  • Research Article
  • Citations440

MRI-Based Attenuation Correction for PET/MRI Using Ultrashort Echo Time Sequences

  • May 01, 2010
  • Journal of Nuclear Medicine
  • Vincent Keereman +5
  • Research Article
  • Citations109

The effect of limited MR field of view in MR/PET attenuation correction

  • May 20, 2010
  • Medical Physics
  • Gaspar Delso +4
  • Research Article
  • Citations25

Surveillance of unruptured intracranial saccular aneurysms using non-contrast 3D black blood MRI: comparison of 3D TOF and CE-MRA with 3D DSA

  • May 23, 2019
  • AJNR. American journal of neuroradiology
  • C Zhu +8
  • Research Article
  • Citations13

Respiratory- and cardiac motion-corrected simultaneous whole-heart PET and dual phase coronary MR angiography.

  • Oct 15, 2018
  • Magnetic resonance in medicine
  • Camila Munoz +5
  • Research Article
  • Citations28

Beat-to-beat respiratory motion correction with near 100% efficiency: a quantitative assessment using high-resolution coronary artery imaging

  • Feb 02, 2011
  • Magnetic Resonance Imaging
  • Andrew D Scott +2
  • PDF
  • Abstract

Improved data acquisition efficiency for respiratory motion correction in coronary MRI

  • Feb 01, 2012
  • Journal of Cardiovascular Magnetic Resonance
  • Mehdi H Moghari +9
Cactus Communications logo

Copyright 2026 Cactus Communications. All rights reserved.