• Home
  • Search
  • Comparison of Gradation Characteristics between Three Typical Medical Imaging Display Devices Using Psychophysical Analysis
  • Cite Icon1
  • https://doi.org/10.6009/jjrt.kj00004022968Copy DOI Icon

Comparison of Gradation Characteristics between Three Typical Medical Imaging Display Devices Using Psychophysical Analysis

Show More
  • Abstract
  • Literature Map
  • References
  • Citations
  • Similar Papers
Abstract

Although there are remarkable differences in maximum luminance in cathode ray tube (CRT) and liquid crystal display (LCD) monitors and film/viewer systems, these differences cannot be recognized in our perception of them. To clarify the reason for this conflict, we analyzed the psychophysical gradient (delta), which is based on the minimum perceptible luminance difference (DeltaL(min)) and can express contrast visually recognized by observers. In this study, we first confirmed the compatibility of the psychophysical analysis to the CRT and the LCD monitors by using their threshold contrasts (C(t)s). Second, we calculated and compared the delta's of the above output devices. The C(t)s values of each device were in good agreement. Moreover, the Moon & Spencer model, which expressed the perceptibility of luminance change, was well suited to the measured C(t)s over the whole luminance range. The psychophysical analysis is therefore available not only for the film/viewer system, but also for the CRT and LCD systems. The difference of physical gradient G of the luminance characteristics curve among the output devices was larger than 20 times, whereas that of d was within 3 times. The display devices listed in the order of decreasing delta were film/viewer>LCD>CRT. These results corresponded to the visual contrast sensation and our clinical experience, which cannot recognize remarkable differences in perception. By using the psychophysical analysis, we clarified the reason for the conflict between the results of physical evaluation and the contrast visually recognized by observers.

Similar Papers
  • Research Article

The TVs of Tomorrow: How RCA's Flat-screen Dreams Led to the First LCDs by Benjamin Gross

  • Jan 01, 2020
  • Technology and Culture
  • Michael Aaron Dennis
  • Research Article

A Comparison between Liquid-Crystal Display Monitors and Cathode-Ray Tube Monitors: A Combined Assessment Study of Observer Performances by Using Storage Phosphor and Flat-Panel-Detector Radiography in Detecting Experimentally Induced Pulmonary Edema in

  • Dec 01, 2009
  • Journal of Medicine and Life Science
  • Sun Young Jeong +5
  • Research Article
  • Citations7

Chromaticity and correlated color temperature of the white point in medical liquid–crystal display

  • Jul 27, 2012
  • Medical Physics
  • Hiroshi Akamine +5
  • PDF
  • Research Article
  • Citations15

How Monitor Characteristics Affect Human Perception in Visual Computer Experiments: CRT vs. LCD Monitors in Millisecond Precise Timing Research

  • Apr 24, 2020
  • Scientific Reports
  • Michaela Rohr +1
  • Research Article
  • Citations9

Comparison of cathode ray tube and liquid crystal display stimulators for use in multifocal VEP.

  • Jul 02, 2014
  • Documenta ophthalmologica. Advances in ophthalmology
  • Marÿke Fox +3
  • Research Article
  • Citations2

OLED: New Generation Display Technology

  • Mar 08, 2022
  • Journal of Ravishankar University (PART-B)
  • Lokesh Kumar Sahu +1
  • Research Article
  • Citations1

Influence of 3D printing parameters on surface roughness and stainability of a denture base resin: An in vitro study.

  • Aug 01, 2025
  • The Journal of prosthetic dentistry
  • Hamile Emanuella Do Carmo Viotto +5
  • Research Article
  • Citations45

Technological and Psychophysical Considerations for Digital Mammographic Displays

  • Mar 01, 2005
  • RadioGraphics
  • Ehsan Samei
  • Research Article
  • Citations7

Grayscale calibration and quality assurance of diagnostic monitors in a PACS system

  • Sep 01, 2006
  • La radiologia medica
  • A Crespi +3
  • Conference Article
  • Citations14

<title>Comparative evaluation of color characterization and gamut of LCD versus CRTs</title>

  • Dec 28, 2001
  • Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE
  • Gaurav Sharma
  • Research Article
  • Citations2

Potential Use of Waste Cathode Ray Tube (CRT) Glass as Fine Aggregate in Concrete for Sustainable Built Environment: A Review

  • Apr 30, 2020
  • International Journal for Research in Engineering Application & Management
  • Shimol Philip +99
  • Research Article
  • Citations5

Noise estimation and reduction on five medical liquid-crystal displays

  • Jan 01, 2006
  • Journal of the Society for Information Display
  • Jiahua Fan +3
  • Research Article
  • Citations48

Achieving precise display timing in visual neuroscience experiments

  • Jun 30, 2010
  • Journal of Neuroscience Methods
  • Tobias Elze
  • Book Chapter

05 - Display Devices

  • Jan 01, 2006
  • High Dynamic Range Imaging
  • Erik Reinhard +3
  • Research Article
  • Citations25

Liquid-Crystal Display Monitors and Cathode-Ray Tube Monitors: A Comparison of Observer Performance in the Detection of Small Solitary Pulmonary Nodules

  • Jan 01, 2003
  • Korean Journal of Radiology
  • Soon-A Hwang +9
Cactus Communications logo