In vitro characterization of PrismaLung+: a novel ECCO2R device

Show More
  • Abstract
  • Highlights & Summary
  • PDF
  • Literature Map
  • References
  • Citations
  • Similar Papers
Abstract

BackgroundInvasive mechanical ventilation is lifesaving in the setting of severe acute respiratory failure but can cause ventilation-induced lung injury. Advances in extracorporeal CO2 removal (ECCO2R) technologies may facilitate more protective lung ventilation in acute respiratory distress syndrome, and enable earlier weaning and/or avoid invasive mechanical ventilation entirely in chronic obstructive pulmonary disease exacerbations. We evaluated the in vitro CO2 removal capacity of the novel PrismaLung+ ECCO2R device compared with two existing gas exchangers.MethodsThe in vitro CO2 removal capacity of the PrismaLung+ (surface area 0.8 m2, Baxter) was compared with the PrismaLung (surface area 0.35 m2, Baxter) and A.L.ONE (surface area 1.35 m2, Eurosets) devices, using a closed-loop bovine blood–perfused extracorporeal circuit. The efficacy of each device was measured at varying pCO2 inlet (pinCO2) levels (45, 60, and 80 mmHg) and blood flow rates (QB) of 200–450 mL/min; the PrismaLung+ and A.L.ONE devices were also tested at a QB of 600 mL/min. The amount of CO2 removed by each device was assessed by measurement of the CO2 infused to maintain circuit equilibrium (CO2 infusion method) and compared with measured CO2 concentrations in the inlet and outlet of the CO2 removal device (blood gas analysis method).ResultsThe PrismaLung+ device performed similarly to the A.L.ONE device, with both devices demonstrating CO2 removal rates ~ 50% greater than the PrismaLung device. CO2 removal rates were 73 ± 4.0, 44 ± 2.5, and 72 ± 1.9 mL/min, for PrismaLung+, PrismaLung, and A.L.ONE, respectively, at QB 300 mL/min and pinCO2 45 mmHg. A Bland–Altman plot demonstrated that the CO2 infusion method was comparable to the blood gas analysis method for calculating CO2 removal. The resistance to blood flow across the test device, as measured by pressure drop, varied as a function of blood flow rate, and was greatest for PrismaLung and lowest for the A.L.ONE device.ConclusionsThe newly developed PrismaLung+ performed more effectively than PrismaLung, with performance of CO2 removal comparable to A.L.ONE at the flow rates tested, despite the smaller membrane surface area of PrismaLung+ versus A.L.ONE. Clinical testing of PrismaLung+ is warranted to further characterize its performance.

Loading PDF

Similar Papers
  • Research Article
  • Citations9

Effect of Hematocrit on the CO2 Removal Rate of Artificial Lungs.

  • Mar 03, 2020
  • ASAIO Journal
  • Alexandra G May +3
  • PDF
  • Research Article
  • Citations8

Extracorporeal Carbon Dioxide Removal: From Pathophysiology to Clinical Applications; Focus on Combined Continuous Renal Replacement Therapy.

  • Jan 05, 2023
  • Biomedicines
  • Francesca Cappadona +9
  • PDF
  • Research Article
  • Citations4

A mock circulation loop to test extracorporeal CO2 elimination setups

  • Sep 11, 2020
  • Intensive Care Medicine Experimental
  • Leonie S Schwärzel +9
  • PDF
  • Research Article
  • Citations22

An extracorporeal carbon dioxide removal (ECCO2R) device operating at hemodialysis blood flow rates

  • Sep 06, 2017
  • Intensive Care Medicine Experimental
  • R Garrett Jeffries +3
  • Research Article

Prediction of Long-Term Mortality in Acute Hypercapnic Respiratory Failure with Use of Low-Flow Veno-Venous Extracorporeal CO2 Removal: A Retrospective Single-Center Study

  • Jan 29, 2026
  • Blood Purification
  • Anne-Aylin Sigg +7
  • Research Article
  • Citations9

The Effect of Pumpless Extracorporeal CO2 Removal on Regional Perfusion of the Brain in Experimental Acute Lung Injury

  • Jul 01, 2013
  • Journal of Neurosurgical Anesthesiology
  • Stefan Kreyer +8
  • Research Article
  • Citations6

Feasibility of Continuous CO2 Removal with Hydrophilic Membranes at Low Blood Flow Rates

  • Mar 01, 2005
  • The International Journal of Artificial Organs
  • P Czermak +3
  • Research Article
  • Citations12

Extracorporeal CO2 Elimination (ECCO2R) for Hypercapnic Respiratory Failure: From Pathophysiology to Clinical Application

  • Apr 13, 2017
  • Pneumologie (Stuttgart, Germany)
  • C Karagiannidis +5
  • Research Article
  • Citations37

Femoral arteriovenous extracorporeal carbon dioxide elimination using low blood flow.

  • Jun 01, 1992
  • Critical Care Medicine
  • J Duncan Young +3
  • Research Article
  • Citations15

In-vitro performance of a low flow extracorporeal carbon dioxide removal circuit.

  • Aug 23, 2019
  • Perfusion
  • Nicholas A Barrett +2
  • Conference Article

Effects of veno-venous extracorporeal CO2 removal on pulmonary hypertension in an ARDS model

  • Sep 01, 2015
  • Philippe Morimont +4
  • Research Article
  • Citations3

Effects of Transglutaminase 2 Inhibition on Ventilator-Induced Lung Injury

  • Jan 01, 2014
  • Journal of Korean Medical Science
  • In Bum Suh +13
  • PDF
  • Research Article
  • Citations33

Control of respiratory drive by extracorporeal CO2 removal in acute exacerbation of COPD breathing on non-invasive NAVA

  • Apr 23, 2019
  • Critical Care
  • Christian Karagiannidis +7
  • Research Article

Extracorporeal Membrane Oxygenation in Acute Respiratory Failure due to Hemorrhagic Alveolitis in a Patient with Acute Myeloblastic Leukemia.

  • Jan 01, 2024
  • Case Reports in Critical Care
  • Martina Cuccarelli +7
  • Research Article
  • Citations27

Initial ventilator settings for critically ill patients

  • Jan 01, 2013
  • Critical Care
  • Oguz Kilickaya +1
Cactus Communications logo

Copyright 2026 Cactus Communications. All rights reserved.