Low-dose lung CT: Optimizing diagnostic radiation dose – A phantom study

Show More
  • Abstract
  • Literature Map
  • References
  • Citations
  • Similar Papers
Abstract

Background/purposeTo investigate a quantitative method for assessing image quality of low dose lung computed tomography (CT) and find the lowest exposure dose providing diagnostic images. MethodsAxial volumetric lung CT acquisitions (256 slice scanner) were performed on three different sized anthropomorphic phantoms at different dose levels. The maximum steepness of sigmoid curves fitted to line density profiles was measured at lung-to-pleura interfaces. For each phantom, image sharpness was calculated as the median of 468 measurements from 39 different locations. Diagnostic image quality for the adult and paediatric phantom was rated by three radiologists using 4-point Likert scales. The image sharpness cut-off for obtaining adequate image quality was determined from qualitative ratings. ResultsAdequate diagnostic image quality was reached at a median steepness of 713 HU/mm in the adult phantom with a corresponding CTDIvol of 0.14 mGy and an effective dose of 0.13 mSv at a dose level of 100 kVp and 10 mA. In the paediatric phantom diagnostic image quality was reached at a median steepness of 1139 HU/mm with a corresponding CTDIvol of 0.13 mGy and an effective dose of 0.08 mSv at a dose level of 100 kVp and 10 mA. ConclusionsDetermination of image sharpness on line density profiles can be used as quantitative measure for image quality of lung CT. Sufficient-quality lung CT can be achieved at effective radiation doses of 0.13 mSv (adult phantom) and 0.08 mSv (paediatric phantom). These findings suggest that substantial dose reduction is feasible without compromising diagnostic accuracy.

Similar Papers
  • PDF
  • Research Article

Attenuation-based ultra-low-dose lung computed tomography at 0.1 mSv to 0.3mSv effective dose in children.

  • Jan 19, 2026
  • Pediatric radiology
  • Max-Johann Sturm +4
  • Research Article

116 IMAGE QUALITY AND RADIATION DOSE WITH SINGLE HEART BEAT 320 MULTIDETECTOR CT CORONARY ANGIOGRAPHY

  • May 01, 2013
  • Heart
  • M C Williams +9
  • Research Article
  • Citations6

Diagnostic image quality of hand-held and wall-mounted X-ray devices in bitewing radiography: a non-inferiority clinical trial.

  • Jan 08, 2021
  • Dento maxillo facial radiology
  • Reinier Cornelis Hoogeveen +2
  • Research Article
  • Citations23

Prospective ECG-triggered sequential scan protocol for coronary dual-source CT angiography: initial experience

  • Jan 28, 2009
  • The International Journal of Cardiovascular Imaging
  • Franziska Hein +8
  • Research Article
  • Citations16

MO‐A‐ValB‐01: Tradeoffs in Image Quality and Radiation Dose for CT

  • Jun 01, 2006
  • Medical Physics
  • M Mcnitt‐Gray
  • Research Article

Image quality assessment and automation in late gadolinium-enhanced MRI of the left atrium in atrial fibrillation patients.

  • Dec 21, 2024
  • Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing
  • Benjamin Orkild +10
  • Research Article
  • Citations220

Impact of a Reduced Tube Voltage on CT Angiography and Radiation Dose: Results of the PROTECTION I Study

  • Aug 01, 2009
  • JACC: Cardiovascular Imaging
  • Bernhard Bischoff +6
  • Research Article
  • Citations20

IMAGING DOSE OF HUMAN ORGANS FROM kV-CBCT IN IMAGE-GUIDED RADIATION THERAPY.

  • Oct 20, 2016
  • Radiation Protection Dosimetry
  • Kihong Son +3
  • Research Article
  • Citations12

Dual-energy CT angiography-derived virtual non-contrast images for follow-up of patients with surgically clipped aneurysms: a retrospective study.

  • Jan 25, 2019
  • Neuroradiology
  • Su Young Yun +9
  • Research Article

Computed tomography radiation dose optimization for the diagnosis of acute rib fractures in infants investigated for suspected physical abuse: An experimental animal study.

  • Nov 01, 2025
  • Child abuse & neglect
  • Nasser M Alzahrani +4
  • Research Article

Feasibility study of coronary CT angiography in single cardiac cycle in patients with high heart rate using 256-row detector CT

  • Feb 10, 2017
  • Chinese journal of radiology
  • Jiantao Liang +4
  • Research Article
  • Citations20

Image Quality of Coronary Arteries on Non-electrocardiography-gated High-Pitch Dual-Source Computed Tomography in Children with Congenital Heart Disease.

  • Jul 08, 2017
  • Pediatric Cardiology
  • Yuichiro Kanie +3
  • Research Article

Randomised trial on pain reduction in hysterosalpingography - a modified technique without vulsellum forceps

  • Dec 31, 2022
  • The ASEAN Journal of Radiology
  • Olubukola Omidiji +7
  • Research Article
  • Citations6

Radiation dose considerations in digital radiography with an anti‐scatter grid: A study using adult and pediatric phantoms

  • Jul 25, 2023
  • Journal of Applied Clinical Medical Physics
  • Hiroki Kawashima +6
  • Research Article
  • Citations15

A projection image database to investigate factors affecting image quality in weight-based dosing: application to pediatric renal SPECT

  • Jul 01, 2018
  • Physics in Medicine & Biology
  • Ye Li +10
Cactus Communications logo

Copyright 2026 Cactus Communications. All rights reserved.