Improvement of Beam-Quality Evaluation Method for Medical Linear Accelerator Using Magnetic Field
- Jun 30, 2015
- Journal of Magnetics
- Jeongho Kim + 4 more +4
Beam-quality of medical linac evaluations vary by diverse factors. Because conventional beam-quality evaluation methods yield fragmentary results, a new beam-evaluation method is suggested, and its feasibility is evaluated. The PDDs (percentage depth doses) of 6 MV (Mega-voltage) and 10 MV photon, R (Range) of a 6 MeV (Mega Electron-voltage) and 9 MeV electron were measured and compared with the conventional evaluation methods, and the improved methods PDD 10 5 , PDD 20 10 , PDD 30 20 , PDD 20 5 , PDD 30 10 , and R 70 50 , R 50 30 , R 70 30 as the magnetic field of the bending magnet was changed to +2% to ?2%, and the results were compared. The comparison showed that the improved methods exhibit a higher discrimination than the conventional methods in each energy regime. PDD 10 5 , PDD 30 20 , PDD 30 10 and R 70 50 , R 50 30 should be applied. These methods exhibit a higher discrimination in each energy regime than conventional beam-quality evaluation methods; therefore, they should be used for beam-quality evaluation according to the magnetic field variation.