  • Noise correlation in PET, CT, SPECT and PET/CT data evaluated using autocorrelation function: a phantom study on data, reconstructed using FBP and OSEM
  • Cite Icon72
  • https://doi.org/10.1186/1471-2342-5-5Copy DOI Icon

Noise correlation in PET, CT, SPECT and PET/CT data evaluated using autocorrelation function: a phantom study on data, reconstructed using FBP and OSEM

Show More
  • Abstract
  • Highlights & Summary
  • PDF
  • Literature Map
  • References
  • Citations
  • Similar Papers
Abstract

BackgroundPositron Emission Tomography (PET), Computed Tomography (CT), PET/CT and Single Photon Emission Tomography (SPECT) are non-invasive imaging tools used for creating two dimensional (2D) cross section images of three dimensional (3D) objects. PET and SPECT have the potential of providing functional or biochemical information by measuring distribution and kinetics of radiolabelled molecules, whereas CT visualizes X-ray density in tissues in the body. PET/CT provides fused images representing both functional and anatomical information with better precision in localization than PET alone.Images generated by these types of techniques are generally noisy, thereby impairing the imaging potential and affecting the precision in quantitative values derived from the images. It is crucial to explore and understand the properties of noise in these imaging techniques. Here we used autocorrelation function (ACF) specifically to describe noise correlation and its non-isotropic behaviour in experimentally generated images of PET, CT, PET/CT and SPECT.MethodsExperiments were performed using phantoms with different shapes. In PET and PET/CT studies, data were acquired in 2D acquisition mode and reconstructed by both analytical filter back projection (FBP) and iterative, ordered subsets expectation maximisation (OSEM) methods. In the PET/CT studies, different magnitudes of X-ray dose in the transmission were employed by using different mA settings for the X-ray tube. In the CT studies, data were acquired using different slice thickness with and without applied dose reduction function and the images were reconstructed by FBP. SPECT studies were performed in 2D, reconstructed using FBP and OSEM, using post 3D filtering. ACF images were generated from the primary images, and profiles across the ACF images were used to describe the noise correlation in different directions. The variance of noise across the images was visualised as images and with profiles across these images.ResultsThe most important finding was that the pattern of noise correlation is rotation symmetric or isotropic, independent of object shape in PET and PET/CT images reconstructed using the iterative method. This is, however, not the case in FBP images when the shape of phantom is not circular. Also CT images reconstructed using FBP show the same non-isotropic pattern independent of slice thickness and utilization of care dose function. SPECT images show an isotropic correlation of the noise independent of object shape or applied reconstruction algorithm. Noise in PET/CT images was identical independent of the applied X-ray dose in the transmission part (CT), indicating that the noise from transmission with the applied doses does not propagate into the PET images showing that the noise from the emission part is dominant. The results indicate that in human studies it is possible to utilize a low dose in transmission part while maintaining the noise behaviour and the quality of the images.ConclusionThe combined effect of noise correlation for asymmetric objects and a varying noise variance across the image field significantly complicates the interpretation of the images when statistical methods are used, such as with statistical estimates of precision in average values, use of statistical parametric mapping methods and principal component analysis. Hence it is recommended that iterative reconstruction methods are used for such applications. However, it is possible to calculate the noise analytically in images reconstructed by FBP, while it is not possible to do the same calculation in images reconstructed by iterative methods. Therefore for performing statistical methods of analysis which depend on knowing the noise, FBP would be preferred.

Loading PDF

Similar Papers
  • Research Article

Effects of the different PET image reconstruction methods on distribution of dopamine transporter in healthy human brain

  • Sep 25, 2018
  • Int J Radiat Med Nucl Med
  • Yuhua Zhu +9
  • Research Article
  • Citations3

Evaluation of Image Reconstruction Techniques in Single Photon Emission Computed Tomography (SPECT/CT) Imaging

  • Apr 20, 2018
  • Research & Reviews: Journal of Physics
  • F Usman +3
  • Research Article

Reducing Artifacts in Pelvic Bone SPECT: An Assessment of Lesion Detectability Using Numerical and Human Observers

  • Oct 01, 2006
  • IEEE Transactions on Nuclear Science
  • T.H Farncombe +2
  • Research Article
  • Citations39

Performance measurement of PSF modeling reconstruction (True X) on Siemens Biograph TruePoint TrueV PET/CT

  • Feb 07, 2014
  • Annals of Nuclear Medicine
  • Young Sub Lee +5
  • Research Article
  • Citations13

Effects of attenuation correction and reconstruction method on PET activation studies

  • Sep 17, 2003
  • NeuroImage
  • Catalina T Mesina +6
  • Research Article
  • Citations21

Radiation Dose Reduction in Pediatric Body CT Using Iterative Reconstruction and a Novel Image-Based Denoising Method.

  • Nov 01, 2015
  • American Journal of Roentgenology
  • Lifeng Yu +6
  • Research Article
  • Citations12

Texture analysis for automated evaluation of Jaszczak phantom SPECT system tests.

  • Dec 06, 2018
  • Medical Physics
  • Kenneth J Nichols +2
  • Research Article
  • Citations11

A detector’s eye view (DEV)-based OSEM algorithm for benchtop x-ray fluorescence computed tomography (XFCT) image reconstruction

  • Apr 01, 2019
  • Physics in Medicine & Biology
  • Luzhen Deng +5
  • Research Article
  • Citations15

The Value of Attenuation Correction in Hybrid Cardiac SPECT/CT on Inferior Wall According to Body Mass Index

  • Jan 01, 2016
  • World Journal of Nuclear Medicine
  • Muge Tamam +3
  • Conference Article
  • Citations5

A mathematical observer study for evaluation of a model-based compensation method for crosstalk in simultaneous dual isotope SPECT

  • Dec 01, 2003
  • X Song +4
  • Research Article
  • Citations37

Impact of Acquisition Geometry, Image Processing, and Patient Size on Lesion Detection in Whole-Body 18F-FDG PET

  • Nov 15, 2007
  • Journal of Nuclear Medicine
  • G El Fakhri +5
  • Research Article

逐次近似SPECT再構成法の<sup>123</sup>I-IMP局所脳血流量定量への応用─最適な再構成法と前処理フィルタの遮断周波数に関する検討─

  • Jan 01, 2013
  • Japanese Journal of Radiological Technology
  • Akira Asazu +6
  • Abstract

328. The Basedow’s disease radiometabolic therapy: The uptake volume estimation

  • Dec 01, 2018
  • Physica Medica
  • A Ostinelli +9
  • Research Article
  • Citations49

Characterization of ordered-subsets expectation maximization with 3d post-reconstruction gauss filtering and comparison with filtered backprojection in99mTc SPECT

  • Apr 01, 2005
  • Annals of Nuclear Medicine
  • Marco Brambilla +5
  • Research Article
  • Citations79

Influence of OSEM and segmented attenuation correction in the calculation of standardised uptake values for [18F]FDG PET

  • Jun 27, 2001
  • European Journal of Nuclear Medicine
  • D Visvikis +5
Cactus Communications logo

Copyright 2026 Cactus Communications. All rights reserved.