Abstract

Background: Despite the development of iterative reconstruction (IR) in diagnostic imaging, CBCT are generally reconstructed with filtered back projection (FBP) in radiotherapy. Varian medical systems, recently released with their latest Halcyon® V2.0 accelerator, a new IR algorithm for CBCT reconstruction. Purpose: To assess the image quality of radiotherapy CBCT images reconstructed with FBP and an IR algorithm. Methods: Three CBCT acquisition modes (head, thorax and pelvis large) available on a Halcyon® were assessed. Five acquisitions were performed for all modes on an image quality phantom and reconstructed with FBP and IR. Task-based image quality assessment was performed with noise power spectrum (NPS), task-based transfer function (TTF) and detectability index (d’). To illustrate the image quality obtained with both reconstruction types, CBCT acquisitions were made on 6 patients. Results: The noise magnitude and the spatial frequency of the NPS peak was lower with IR than with FBP for all modes. For all low and high-contrast inserts, the values for TTF at 50% were higher with IR than with FBP. For all inserts and all modes, the contrast values were similar with FBP and IR. For all low and high-contrast simulated lesions, d’ values were higher with IR than with FBP for all modes. These results were also found on the 6 patients where the images were less noisy but smoother with IR-CBCT. Conclusions: Using the IR algorithm for CBCT images in radiotherapy improve image quality and thus could increase the accuracy of online registration and limit positioning errors during processing.

Loading PDF

Similar Papers
  • Research Article
  • Citations127

Six iterative reconstruction algorithms in brain CT: a phantom study on image quality at different radiation dose levels

  • Sep 18, 2013
  • The British Journal of Radiology
  • A Löve +5
  • Research Article
  • Citations26

The feasibility of sub-millisievert coronary CT angiography with low tube voltage, prospective ECG gating, and a knowledge-based iterative model reconstruction algorithm.

  • Oct 31, 2015
  • The International Journal of Cardiovascular Imaging
  • Chul Hwan Park +4
  • Research Article
  • Citations14

Iterative CT reconstruction in abdominal low-dose CT used for hybrid SPECT-CT applications: effect on image quality, image noise, detectability, and reader’s confidence

  • Jun 01, 2019
  • Acta Radiologica Open
  • Oliver S Grosser +11
  • Research Article
  • Citations185

Evaluation of Heavily Calcified Vessels with Coronary CT Angiography: Comparison of Iterative and Filtered Back Projection Image Reconstruction

  • Jun 21, 2011
  • Radiology
  • Matthias Renker +10
  • Research Article
  • Citations51

Stenosis Quantification in Coronary CT Angiography

  • Jan 01, 2013
  • Investigative Radiology
  • Fabian Morsbach +7
  • Research Article

TU‐D‐207A‐00: Quantitative Assessment of CT Systems with Iterative Image Reconstruction Algorithms

  • Jun 01, 2016
  • Medical Physics
  • Guang‐Hong Chen +1
  • Research Article
  • Citations5

Technical Note: Scanner dependence of adaptive statistical iterative reconstruction with 3D noise power spectrum central frequency and noise magnitude ratios.

  • Jul 31, 2021
  • Medical Physics
  • Akira Hasegawa +3
  • Research Article
  • Citations6

The Acceptability of Iterative Reconstruction Algorithms in Head CT: An Assessment of Sinogram Affirmed Iterative Reconstruction (SAFIRE) vs. Filtered Back Projection (FBP) Using Phantoms

  • May 31, 2017
  • Journal of Medical Imaging and Radiation Sciences
  • Martine A Harris +3
  • Research Article
  • Citations41

Radiation dose reduction in cerebral CT perfusion imaging using iterative reconstruction

  • Oct 15, 2013
  • European Radiology
  • Joris M Niesten +8
  • Research Article
  • Citations72

Impact of iterative reconstruction on CNR and SNR in dynamic myocardial perfusion imaging in an animal model

  • Jul 03, 2012
  • European Radiology
  • B M Gramer +9
  • PDF
  • Research Article
  • Citations12

Feasibility of radiation dose reduction with iterative reconstruction in abdominopelvic CT for patients with inappropriate arm positioning

  • Dec 31, 2018
  • PLoS ONE
  • Nieun Seo +5
  • Research Article

Phantom-based evaluation of image quality in Transformer-enhanced 2048-matrix CT imaging at low and ultralow doses.

  • Apr 07, 2025
  • Japanese journal of radiology
  • Qingyao Li +8
  • PDF
  • Research Article
  • Citations19

Performance evaluation of two iterative reconstruction algorithms, MBIR and ASIR, in low radiation dose and low contrast dose abdominal CT in children.

  • Apr 21, 2020
  • La radiologia medica
  • Jihang Sun +6
  • Research Article
  • Citations47

Systematic Error in Lung Nodule Volumetry: Effect of Iterative Reconstruction Versus Filtered Back Projection at Different CT Parameters

  • Dec 01, 2012
  • American Journal of Roentgenology
  • Martin J Willemink +7
  • Research Article
  • Citations2

Coronary CT Angiography with Use of Iterative Reconstruction Algorithm in Coronary Stenting: A Systematic Review of Image Quality, Diagnostic Value and Radiation Dose

  • Feb 01, 2015
  • Journal of Medical Imaging and Health Informatics
  • Ahmed Al Shammakhi +1
Cactus Communications logo

Copyright 2026 Cactus Communications. All rights reserved.