- Research Article
- 10.1109/ojemb.2025.3548838
Assessing the Accuracy of Bed-Occupancy With a tina.care Bed Sensor in Hospital Wards and Home Care Settings: A Pilot Study
- Jan 01, 2025
- IEEE Open Journal of Engineering in Medicine and Biology
- Tomáš Kulhánek + 3 more +3
<i>Goal:</i> This pilot study aims to assess accuracy in detecting patient presence or absence by using a bed sensor based on mmwave radar technology above the patient bed. <i>Methods:</i> Patients and healthy volunteers were observed during their presence or absence in a bed in hospital and home location. These observations were compared with data coming from bed sensor monitoring patient presence using tina.care bed sensor ASWA. <i>Results:</i> A total of 53 different observations were performed during the study period and the bed sensor reached accuracy of 94%, precision of 90%, sensitivity of 99% and specificity of 89% to detect presence or absence of patients in a bed. <i>Conclusions:</i> The sensor demonstrated strong performance in detecting patient presence in bed, with reasonable specificity and low false negatives. Further research should assess bed-exit and bed-entry events, system's accuracy in a larger cohort, its impact on patient care, and the precision of vital health parameters measured by the sensor in order to compare it with similar studies.
Read more