Preclinical Evaluation of a Novel Single-Shot Pulsed Field Ablation System for Pulmonary Vein and Atrial Ablation
Abstract BACKGROUNDPulsed field ablation (PFA) is a non-thermal ablative strategy that achieves cell death via electroporation. We investigated the preclinical safety and efficacy of PFA using two novel 8-French, 16-electrode spiral PFA/mapping catheters (CRC EP, Inc).METHODSBipolar PFA (>1.8 kV) was performed using 30 s, single-shot, QRS-gated applications. Ninety-four atrial structures were ablated in 23 swine, 1 canine, and 1 ovine, including right and left atria and atrial appendages, pulmonary veins, and superior and inferior (IVC) vena cavae. Additionally, we examined the impact of PFA on the phrenic nerve in 14 swine and on a deviated esophagus during ablation from inside the IVC in 5 swine.RESULTSAll applications were single-shot without repositioning the catheter. Minimal microbubbling was observed without significant skeletal muscle twitching/activation (mean acceleration: 0.05 m/s2). There was marked reduction in the post-versus pre-PFA atrial electrogram amplitude (0.17±0.21 mVvs. 1.18±1.08 mV; P<0.0001). Durable conduction block was demonstrated up to 3 months in all targeted tissues. Lesions were contiguous and transmural, measuring 25±9 mm x 21±7 mm without any thermal effects. Magnetic resonance, gross, and histologic examinations of the brain, rete mirabile and kidneys revealed no thromboembolism. No acute/long-term phrenic nerve dysfunction was encountered. Though within 2 hours of ablation, histologic examinations of the esophagus revealed acute PFA-related changes in the muscular layer, these completely resolved by 21±5 days.CONCLUSIONA novel, single-shot, spiral PFA system is capable of safely creating large, durable atrial lesions without significant adverse effects on the phrenic nerve or the esophagus.
Read more