B-067 The Challenges in the Validation of Kappa and Lambda Free Light Chain Assays on an Automated Analyzer
Abstract Background Kapp and lambda, serum free light chain (FLC) measurements are pivotal analytes for the diagnosis and monitoring of monoclonal gammopathy or AL amyloidosis. The calculated FLC ratio (kappa/lambda) is strongly recommended for disease diagnosis, risk stratification, and management of plasma dyscrasias. UConn Health Center Laboratory Medicine validated Diazyme’s human kappa and lambda FLC assays to meet the need of quantifying serum FLC in patients suspected of monoclonal gammopathies. UConn Laboratory utilized Abbott’s Alinity c as an automated platform to validate Diazyme’s assay in-house, in place of reference laboratory’s Optilite® Freelite, a quantitative immunoturbidometric assay. Methods Within-run precision, between-run precision, method comparison, and reference range were conducted using CLSI guidelines. For within-run precision, QC levels 1 and 2 were run 10 times each while between-run precision was carried out for 20 days. Method comparison included quantitative evaluation by Deming regression between Optilite® Freelite and Diazyme kappa/lambda assay. For the reference range study, 26 samples were originally obtained for verification of the manufacturer suggested reference range. This eventually turned into establishing a reference range using 122 samples. Results The linearity of kappa FLC (KFLC)assay was measured between 0.00 to 191.20 mg/L, with lambda FLC (LFLC) over a measured range of 0.03 to 213.90 mg/L. KFLC level 1 was found to have a mean of 16.47 mg/L, with a CV 4.2%. KFLC level 2 mean 28.41 mg/dL, with a CV 1.8%. LFLC level 1 mean 24.48 mg/dL, with a CV 2.1%. LFLC level 2 mean of 42.40 mg/dL , CV 1.7%. Quantitative comparison between Optilite® Freelite and Alinity c Diazyme kappa/lambda assay indicated a -1.423 bias for KFLC and 3.212 bias for LFLC. 26 samples were used to verify Diazyme suggested reference range 2.37-20.73 mg/L for KFLC and 4.23-27.69 mg/dL for LFLC. The reference range passed for LFLC, however, failed for KFLC with 23.1% outside the given reference range. 122 samples were used to establish reference range for KFLC and LFLC. Through transformed parametric, reference range for KFLC and LFLC is 6.8-32.2 mg/L and 4.7-24.3 mg/L respectively. The reference range for k/l ratio is 0.61-2.58. Conclusion The median KFLC in normal population in UConn Health was 15.85 mg/L and median LFLC was 11.95 mg/L with k/l ratio median of 1.42. Other studies have shown the need for establishing local reference range for free light chains. However, the dominance of known reference range of 0.26-1.65 creates a problem for clinicians to adopt local reference range. Many laboratories using Diazyme have adopted Diazyme reference range when switching from Optilite to Diazyme kappa/lambda assay. Our continued discussion with hematology and oncology clinicians have geared us towards adopting Diazyme reference range although the established reference range would be different. For future direction, we will analyze local reference range with patient outcome seen at UConn Health through kappa/lambda measurements to understand if reference range revision is needed.
Read more