Detector evaluation of a prototype amorphous selenium-based full field digital mammography system

  • Abstract
  • Literature Map
  • Citations
  • Similar Papers
Abstract

This study evaluated the physical performance of a selenium-based direct full-field digital mammography prototype detector (Siemens Mammomat NovationDR), including the pixel value vs. exposure linearity, the modulation transfer function (MTF), the normalized noise power spectrum (NNPS), and the detective quantum efficiency (DQE). The current detector is the same model which received an approvable letter from FDA for release to the US market. The results of the current prototype are compared to those of an earlier prototype. Two IEC standard beam qualities (RQA-M2: Mo/Mo, 28 kVp, 2 mm Al; RQA-M4: Mo/Mo, 35 kVp, 2 mm Al) and two additional beam qualities (MW2: W/Rh, 28 kVp, 2 mm Al; MW4: W/Rh, 35 kVp, 2 mm Al) were investigated. To calculate the modulation transfer function (MTF), a 0.1 mm Pt-Ir edge was imaged at each beam quality. Detector pixel values responded linearly against exposure values (R2 0.999). As before, above 6 cycles/mm Mo/Mo MTF was slightly higher along the chest-nipple axis compared to the left-right axis. MTF was comparable to the previously reported prototype, with slightly reduced resolution. The DQE peaks ranged from 0.71 for 3.31 μC/kg (12.83 mR) to 0.4 for 0.48 μC/kg (1.86 mR) at 1.75 cycles/mm for Mo/Mo at 28 kVp. The DQE range for W/Rh at 28 kVP was 0.81 at 2.03 μC/kg (7.87 mR) to 0.50 at 0.50 μC/kg (1.94 mR) at 1 cycle/mm. NNPS tended to increase with greater exposures, while all exposures had a significant low-frequency component. Bloom and detector edge artifacts observed previously were no longer present in this prototype. The new detector shows marked noise improvement, with slightly reduced resolution. There remain artifacts due to imperfect gain calibration, but at a reduced magnitude compared to a prototype detector.

Similar Papers
  • Abstract
  • Citations2

Detective quantum efficiency (DQE) of the Dexela 2923MAM detector according to IEC 62220-1-1:2015

  • Aug 26, 2016
  • Physica Medica
  • V Koukou +10
  • Abstract

Getting to grips with quantitative image quality measurements for QA of digital mammography (DM) systems

  • Jul 01, 2016
  • Physica Medica
  • James Mccullagh +3
  • Research Article
  • Citations96

Physical characterization of a scanning photon counting digital mammography system based on Si‐strip detectors

  • May 08, 2007
  • Medical Physics
  • Magnus Åslund +3
  • Research Article
  • Citations17

A novel method for the optimization of positron emission tomography scanners imaging performance.

  • Nov 08, 2016
  • Hellenic journal of nuclear medicine
  • C Michail +8
  • Research Article
  • Citations3

Cascaded systems analysis of shift‐variant image quality in slit‐scanning breast tomosynthesis

  • Aug 28, 2018
  • Medical Physics
  • Karl Berggren +3
  • Research Article
  • Citations17

Cassette-based digital mammography.

  • Oct 01, 2004
  • Technology in Cancer Research & Treatment
  • J Anthony Seibert +3
  • Research Article
  • Citations1

The detective quantum efficiency (DQE) for evaluating the performance of a small gamma camera system with a uniformly redundant array (URA) collimator

  • Mar 21, 2008
  • Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
  • Hosang Jeon +1
  • Abstract

74. Mean glandular dose survey in the Emilia-Romagna region breast screening program

  • Dec 01, 2018
  • Physica Medica
  • Paola Golinelli +25
  • Research Article
  • Citations37

Quality control measurements for digital x-ray detectors

  • Jan 19, 2011
  • Physics in Medicine & Biology
  • N W Marshall +2
  • Conference Article
  • Citations4

Methodology for the objective assessment of lesion detection performance with breast tomosynthesis and digital mammography using a physical anthropomorphic phantom

  • Mar 09, 2018
  • Medical Imaging 2018: Physics of Medical Imaging
  • Lynda C Ikejimba +7
  • Research Article
  • Citations6

Performance evaluation of contrast-detail in full field digital mammography systems using ideal (Hotelling) observer vs. conventional automated analysis of CDMAM images for quality control of contrast-detail characteristics

  • Feb 28, 2015
  • Physica Medica
  • Ioannis Delakis +3
  • Conference Article
  • Citations19

<title>Physical measures of image quality in photostimulable phosphor radiographic systems</title>

  • May 02, 1997
  • Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE
  • Ehsan Samei +1
  • Abstract

OA158] Noise equivalent number of quanta (NEQ): A tool for choosing the optimal reconstruction algorithm in computed tomography (CT)

  • Aug 01, 2018
  • Physica Medica
  • Vittorio Di Trapani +1
  • Conference Article
  • Citations3

A novel x-ray detector design with higher DQE and reduced aliasing: Theoretical analysis of x-ray reabsoprtion in detector converter material

  • Mar 22, 2016
  • Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE
  • Tomi Nano +3
  • Conference Article
  • Citations49

DQE of direct and indirect digital radiography systems

  • Jun 28, 2001
  • Proceedings of SPIE, the International Society for Optical Engineering/Proceedings of SPIE
  • Ehsan Samei +3
Cactus Communications logo

Copyright 2026 Cactus Communications. All rights reserved.