  • Physical evaluation of an ultra-high-resolution CT scanner
  • Cite Icon93
  • https://doi.org/10.1007/s00330-019-06635-5Copy DOI Icon

Physical evaluation of an ultra-high-resolution CT scanner

Show More
  • Abstract
  • PDF
  • Literature Map
  • References
  • Citations
  • Similar Papers
Abstract

ObjectivesTo evaluate the technical performance of an ultra-high-resolution CT (UHRCT) system.MethodsThe physico-technical capabilities of a novel commercial UHRCT system were assessed and compared with those of a current-generation multi-detector (MDCT) system. The super-high-resolution (SHR) mode of the system uses 0.25 mm (at isocentre) detector elements (dels) in the in-plane and longitudinal directions, while the high-resolution (HR) mode bins two dels in the longitudinal direction. The normal-resolution (NR) mode bins dels 2 × 2, resulting in a del-size equivalent to that of the MDCT system. In general, standard procedures and phantoms were used to perform these assessments.ResultsThe UHRCT MTF (10% MTF 4.1 lp/mm) is twice as high as that of the MDCT (10% MTF 1.9 lp/mm), which is comparable to the MTF in the NR mode (10% MTF 1.7 lp/mm). The width of the slice sensitivity profile in the SHR mode (FWHM 0.45 mm) is about 60% of that of the MDCT (FWHM 0.77 mm). Uniformity and CT numbers are within the expected range. Noise in the high-resolution modes has a higher magnitude and higher frequency components compared with MDCT. Low-contrast visibility is lower for the NR, HR and SHR modes compared with MDCT, but about a 14%, for NR, and 23%, for HR and SHR, dose increase gives the same results.ConclusionsHR and SHR mode scanning results in double the spatial resolution, with about a 23% increase in dose required to achieve the same low-contrast detectability.Key Points• Resolution on UHRCT is up to twice as high as for the tested MDCT.• With abdominal settings, UHRCT needs higher dose for the same low-contrast detectability as MDCT, but dose is still below achievable levels as defined by current diagnostic reference levels.• The UHRCT system used in normal-resolution mode yields comparable resolution and noise characteristics as the MDCT system.

Loading PDF

Similar Papers
  • Research Article
  • Citations4

Performance of high-resolution CT for detection and discrimination tasks related to stenotic lesions - A phantom study using model observers.

  • Jan 12, 2023
  • Medical physics
  • Andrew M Hernandez +4
  • Research Article
  • Citations25

Validation of synthesized normal-resolution image data generated from high-resolution acquisitions on a commercial CT scanner.

  • Aug 05, 2020
  • Medical Physics
  • Andrew M Hernandez +8
  • Research Article
  • Citations7

Quantification of airway dimensions using a high-resolution CT scanner: A phantom study.

  • Aug 18, 2021
  • Medical physics
  • Yixiao Zhao +3
  • Research Article
  • Citations28

Accuracy of linear and volumetric measurements of artificial ERR cavities by using CBCT images obtained at 4 different voxel sizes and measured by using 4 different software: an ex vivo research.

  • Jun 27, 2018
  • Dentomaxillofacial Radiology
  • Gül Sönmez +2
  • Research Article
  • Citations1

RELATIONS OF ANISOTROPIC ELASTIC MODULI TO DENSITY AND CT NUMBER IN BOVINE CORTICAL BONE

  • Jun 01, 2008
  • Biomedical Engineering: Applications, Basis and Communications
  • Jui-Ting Hsu +2
  • Research Article

SU‐G‐TeP2‐11: Initial Evaluation of a Novel Split‐Filter Dual‐Energy CT for Use in Radiation Oncology

  • Jun 01, 2016
  • Medical Physics
  • J Miller +3
  • Research Article

WE-D-204-06: An Open Source ImageJ CatPhan Analysis Tool

  • Jun 01, 2015
  • Medical Physics
  • G Nelson
  • Research Article
  • Citations10

View box luminance measurements and their effect on reader performance

  • Sep 01, 1999
  • Academic Radiology
  • Lynn N Rill +2
  • Research Article
  • Citations106

Scatter correction for full‐fan volumetric CT using a stationary beam blocker in a single full scan

  • Oct 20, 2011
  • Medical Physics
  • Tianye Niu +1
  • Abstract
  • Citations1

National DRLs for adult body CT scans based on body width

  • Aug 15, 2014
  • Physica Medica
  • Dara Murphy +5
  • Research Article
  • Citations1

SU-E-J-21: An Intercomparison of Imaging Performance of Two Linac-Mounted Imaging Systems Used in Radiation Therapy: TrueBeam and Trilogy.

  • Jun 01, 2012
  • Medical physics
  • C Kim +3
  • Research Article
  • Citations27

Dual-Energy Computed Tomography for the Characterization of Intracranial Hemorrhage and Calcification: A Systematic Approach in a Phantom System.

  • Jul 28, 2016
  • Investigative Radiology
  • Jessica L Nute +4
  • Conference Article
  • Citations12

Dual energy iodine contrast CT with monochromatic X-rays

  • Jun 29, 2015
  • F.A Dilmanian +21
  • Research Article
  • Citations3

Resolution enhancement of lung 4D‐CT data using multiscale interphase iterative nonlocal means

  • May 01, 2013
  • Medical Physics
  • Yu Zhang +6
  • Research Article
  • Citations4

Reproducibility and stability of voluntary deep inspiration breath hold and free breath in breast radiotherapy based on real-time 3-dimensional optical surface imaging system

  • Nov 11, 2024
  • Radiation Oncology
  • Junxiang Wu +8
Cactus Communications logo

Copyright 2026 Cactus Communications. All rights reserved.