Reduction of motion artifacts in cardiac CT based on partial angle reconstructions from short scan data

  • Abstract
  • Literature Map
  • References
  • Citations
  • Similar Papers
Abstract

Until today, several software-based approaches to increase the temporal resolution in cardiac computed tomography by estimating motion vector fields (MVFs) have been developed. Thereunder, the majority are motion compensation algorithms, which estimate the MVFs employing a three-dimensional registration routine working on reconstructions of multiple cardiac phases.<sup>2, 6, 7, 12 </sup>We present an algorithm that requires nothing more than the data needed for a short scan reconstruction for motion estimation and motion-compensated reconstruction, which both are based on the reconstruction of volumes from a limited angular range.<sup>2, 3, 7, 8</sup> Those partial angle reconstructions are centered at different time points during the short scan and have a temporal resolution of about 10ms each. The MVFs are estimated by a constrained cost function optimization routine employing a motion artifact measuring cost function. During optimization, the MVFs are applied directly by warping the partial angle reconstructions, and the motion compensation is established by simply adding the shifted images. In order to enforce smooth vector fields and keep the number of parameters low, the motion is modeled by a low degree polynomial. Furthermore, to find a good estimation of the MVFs even in phases with rapid cardiac motion, the constrained optimization is re-initialized multiple times. The algorithm is validated with the help of a simulation study and applied to patient data, where motion- compensated reconstructions are performed in various cardiac phases. We show that the image quality can be improved, also in more rapid cardiac phases due to re-initialization of the optimization routine.

Similar Papers
  • Research Article
  • Citations12

Motion vector field phase-to-amplitude resampling for 4D motion-compensated cone-beam CT

  • Feb 01, 2018
  • Physics in Medicine & Biology
  • Sebastian Sauppe +5
  • Research Article
  • Citations51

Cardiac motion correction based on partial angle reconstructed images in x‐ray CT

  • May 01, 2015
  • Medical Physics
  • Seungeon Kim +2
  • Conference Article
  • Citations1

Image-based deformable motion compensation in cone-beam CT: translation to clinical studies in interventional body radiology

  • Mar 16, 2020
  • Sarah Capostagno +5
  • Research Article
  • Citations3

Motion compensation for aortic valves using partial angle CT reconstructions.

  • Feb 11, 2022
  • Medical Physics
  • Sergej Lebedev +4
  • Research Article
  • Citations68

Motion-compensated iterative cone-beam CT image reconstruction with adapted blobs as basis functions

  • Nov 07, 2008
  • Physics in Medicine & Biology
  • A A Isola +4
  • Research Article
  • Citations184

PROPELLER MRI: clinical testing of a novel technique for quantification and compensation of head motion.

  • Aug 23, 2001
  • Journal of Magnetic Resonance Imaging
  • Kirsten P.N Forbes +3
  • Research Article
  • Citations112

Proposal of a low-dose, long-pitch, dual-source chest CT protocol on third-generation dual-source CT using a tin filter for spectral shaping at 100kVp for CoronaVirus Disease 2019 (COVID-19) patients: a feasibility study.

  • Apr 01, 2020
  • La radiologia medica
  • Andrea Agostini +7
  • Conference Article

Free-breathing High Quality Pulmometry MRI: 3D Cones Acquisition with Motion Compensation

  • Sep 16, 2025
  • Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
  • Bochao Li +6
  • Research Article
  • Citations21

Automatic generation of time resolved motion vector fields of coronary arteries and 4D surface extraction using rotational x-ray angiography

  • Dec 08, 2008
  • Physics in Medicine & Biology
  • Uwe Jandt +3
  • Conference Article
  • Citations1

Mismatch Correction for Free-breathing PET and Deep-inspiration Breath-holding CT in PET/CT Imaging

  • Oct 01, 2017
  • Hongcheng Shi +4
  • Research Article
  • Citations4

A Novel Motion Compensation Method Based on Height Term Removing for Spotlight SAR on Circular Trajectories

  • Mar 01, 2020
  • IEEE Sensors Journal
  • Feng Zuo +5
  • PDF
  • Research Article
  • Citations7

Automatic motion compensation for structured illumination endomicroscopy using a flexible fiber bundle.

  • Feb 25, 2020
  • Journal of Biomedical Optics
  • Andrew D Thrapp +1
  • Research Article
  • Citations3

Motion Detection in Compressed Video Using Macroblock Classification

  • May 31, 2014
  • Advanced Computing: An International Journal
  • M Usha
  • Conference Article
  • Citations7

An efficient non-selective adaptive motion compensated frame rate up conversion

  • May 01, 2017
  • Nguyen Van Thang +1
  • Research Article
  • Citations1

Multi-resolution motion estimation and compensation based on adjacent prediction of frame difference in wavelet domain

  • May 01, 2009
  • Journal of Electronics (China)
  • Guowei Tang +1
Cactus Communications logo

Copyright 2026 Cactus Communications. All rights reserved.